Pacer autosequence names – Fluke Biomedical medTester 5000C User Manual

Page 255

Advertising
background image

Biomedical Test System

SPO2 Module

15

15-27

Print Your Autosequence

From the SPO2 custom autosequences menu described above, printing any
autosequence is accomplished through this menu command:

PRINT

F3

Reinitialize Factory Default Settings

All ten autosequences can be returned to the factory default settings. To do
this, enter these medTester commands:

1.

From the medTester 5000C main menu,

MENU

1, press the

right arrow to access

MENU 2

.

2.

Enter the command:

CUSTOM

F3

3.

From the custom menu, press the right arrow and enter the

command:

INIT

F3

4.

Choose:

SPO2

F5

5.

To initialize SPO2 autosequences to defaults, press:

YES

F5