Pacer autosequence names – Fluke Biomedical medTester 5000C User Manual
Page 255
Advertising
![background image](/files/785336/content/doc255.png)
Biomedical Test System
SPO2 Module
15
15-27
Print Your Autosequence
From the SPO2 custom autosequences menu described above, printing any
autosequence is accomplished through this menu command:
F3
Reinitialize Factory Default Settings
All ten autosequences can be returned to the factory default settings. To do
this, enter these medTester commands:
1.
From the medTester 5000C main menu,
MENU
1, press the
right arrow to access
MENU 2
.
2.
Enter the command:
CUSTOM
F3
3.
From the custom menu, press the right arrow and enter the
command:
INIT
F3
4.
Choose:
SPO2
F5
5.
To initialize SPO2 autosequences to defaults, press:
YES
F5