Ipt-1, Ipt-mc – Fluke Biomedical medTester 5000C User Manual

Page 147

Advertising
background image

Biomedical Test System

IV Pump Module

10

10-9

1.

OP CODE—This is the operator code for the person doing the
testing. This code can consist of a maximum of three characters
which can be entered from the medTester keyboard. An example
of an operator code could be a person’s initials or employee
number. The previous operator’s code, if any, will appear. If the
current operator has not changed from previous use, press the

ENT

key.

2.

TYPE—Enter the device name, code, or other descriptor for the
equipment under test. This entry can be a maximum of 16
characters.

3.

MANF—Enter the name of the device manufacturer, 16
characters maximum.

4.

LOC—Type in the location of the device, 16 characters
maximum. The previous location, if any, will appear. If this occurs
and is the desired location, press the

ENT

key.

5.

MODEL—Enter the device model number, 16 characters
maximum.

6.

SERIAL #—Type the device serial number, again 16 characters
maximum.

7.

CONTROL #—Enter a control number with a maximum of 16
characters.

8.

PHYSICAL INSPECTION—You have two physical inspection
lines, each of which can consist of a maximum of 40 characters.
After entering data for the first prompt, the second prompt
appears.

Note

See Chapter 5, Autosequences, for a complete explanation of
working with prompts. You can turn individual prompts on or
off and rename the prompts numbered six through eight (6-8)
above. Chapter 6, Customize Your medTester 5000C, explains how
to customize prompts.