The wedge adapter, Features of the wedge – Fluke Biomedical medTester 5000C User Manual

Page 193

Advertising
background image

Biomedical Test System

Remote Operation

12

12-15

MEASUREMENT COMMANDS

(µA RMS or DC)

LALD

ECG leakage current, Left Arm to leads.
(µA RMS or DC)

LLLD

ECG leakage current, Left Leg to leads.
(µA RMS or DC)

VLLD

ECG leakage current, V leads to leads.
(µA RMS or DC)

EISO

ECG leakage current, All leads isolation.
(µA RMS or DC)

RLIS

ECG leakage current, Right Leg isolation.
(µA RMS or DC)

RAIS

ECG leakage current, Right Arm isolation.
(µA RMS or DC)

LAIS

ECG leakage current, Left Arm isolation.
(µA RMS or DC)

LLIS

ECG leakage current, Left Leg isolation.
(µA RMS or DC)

VISO

ECG leakage current, V leads isolation.
(µA RMS or DC)

ECUR Equipment

current.

(A RMS)

CRST

Power cord resistance.
(OHMS)

XTMR

Power cord resistance.
(OHMS)

IPWR

Isolated power adapter leakage current.
(µA RMS or DC)

TAC

Tachometer revolutions per minute.
(RPM)

TEMP Temperature.

(DEG F or DEG C)

H2O Humidity.

(%)

REPEAT

Asks for another reading for the measurement which has already been
set up by a previous command. Returns the reading.

EXIT

Turns off a measurement setup or wave which is running. Returns a *.

GTOL

Same as EXIT