Biomedical Test System

Autosequences

5

5-9

Leakage Current Measurements

After the power cord resistance test, the autosequence moves to leakage
current measurements with a message:

Note

This message appears only if the power-off test prompt is turned
on and the power-off tests are enabled. Leakage current tests are
always executed with power on and can not be turned off.

The medTester 5000C now executes all leakage current tests in the
autosequence—first power-off readings (except where power-off tests are
turned off), then power-on readings . To begin the power-off or power-on

tests press

READY

F5

when prompted.

The medTester sets the test receptacle to the correct test configuration. A
five-second delay occurs whenever the test receptacle switches between
normal and reverse polarity. If a

#

(pound) sign appears, an out-of-limit

condition is indicated.

Note

For a complete list and description of leakage current tests, see
Chapter 6, Customize Your medTester 5000C.

Equipment Current Measurement
This test measures the current used by the EUT.

TURN EUT POWER OFF