Isolation lead leakage – Fluke Biomedical medTester 5000C User Manual

Page 49

Advertising
background image

Biomedical Test System

Manual Tests

4

4-3

External Case Leakage

See Figure 4-2, Case External Leakage.

1.

Plug the EUT (equipment under test) into the test receptacle.

2.

Connect the dual banana-plug end of a Kelvin cable to the two
red external meter posts.

3.

Connect the alligator-clip end of the Kelvin cable to a grounded
point on the case of the equipment under test.

4.

Enter menu commands:

CASEX

F1

µA mV

F2

MANUAL

F2

5.

Select the desired condition:

GROUND—closed or open,

POLARITY—normal or reversed,

NEUTRAL—open or closed.

The respective condition

LEDs

light, and the display reads the

current leakage in

µA.

External leakage current is the current that flows from the case of
the equipment under test to earth ground. The current flowing
from the case directly to ground through an external path is
measured.

Internal Case Leakage

See Figure 4-3, Case Internal Leakage.

1.

Plug the EUT (equipment under test) into the test receptacle.

2.

Enter menu commands:

CASEIN

F2

µA mV

F2

MANUAL

F2

3.

Select the desired condition: