What™s a checklist – Fluke Biomedical medTester 5000C User Manual

Page 163

Advertising
background image

Biomedical Test System

IV Pump Module

10

10-25

4.

RATE LOW LIMIT: Enter the low limit for the flow rate. 0000.0
turns off low limit checking.

5.

TEST END TYPE: Select

6.

MANUAL to end the test by pressing a key.

7.

TIMED to end the test after a programmed time:

8.

END TIME: Enter the elapsed time that will end the
test, 00:01 to 99:59 (hours:minutes).

9.

VOLUME to end the test after a specific volume.

10.

END VOLUME: Enter the volume that will end the
test, 0000.1 to 9999.0 ml.

IPT-MC

For the IPT-MC, there is only 1 type of test available, a regular flow test:

1.

FLOW RATE: Enter the nominal flow rate, 0000.1 to 0999.0
ml/hr.

2.

RATE HIGH LIMIT: Enter the high limit for the flow rate.
0000.0 turns off high limit checking.

3.

RATE LOW LIMIT: Enter the low limit for the flow rate. 0000.0
turns off low limit checking.

4.

TEST END TYPE: Select

5.

MANUAL to end the test by pressing a key.

6.

TIMED to end the test after a programmed time:

7.

END TIME: Enter the elapsed time that will end the
test, 00:01 to 99:59 (hours:minutes).

8.

VOLUME to end the test after a specific volume.

9.

END VOLUME: Enter the volume that will end the
test, 0000.1 to 9999.0 ml.

After completing these entries, for up to ten (10) steps, the medTester
prompts you to store the autosequence:

STORE

F5

Print Your Autosequence
From the IV pump autosequences menu described above, printing any
autosequence is accomplished through this menu command: