Lead to lead leakage – Fluke Biomedical medTester 5000C User Manual
Page 48
Advertising
medTester 5000C
Operators Manual
4-2
Hot to Neutral
Enter menu commands:
MANUAL
F2
VOLTS
F1
HOT-NEUT
F1
Neutral to Ground
Enter menu commands:
MANUAL
F2
VOLTS
F1
NEUT-GND
F2
Hot to Ground
Enter menu commands:
HOT-GND
F3
MANUAL
F2
VOLTS
F1
Leakage Current
All leakage current measurements are taken through a 1–k
Ω AAMI load RMS
or DC. Current is measured in a range of 2000
µA or 200 µA.
Note
These tests are described using Kelvin cables for connections.
However, single-wire test cables can be used. Connect single-
wire test cables to the external input posts only.
Note
For all leakage current measurements, except ECG lead
isolation,
DC
toggles the measurement between RMS and DC.