Lead to lead leakage – Fluke Biomedical medTester 5000C User Manual

Page 48

Advertising
background image

medTester 5000C
Operators Manual

4-2

Hot to Neutral

Enter menu commands:

MANUAL

F2

VOLTS

F1

HOT-NEUT

F1

Neutral to Ground

Enter menu commands:

MANUAL

F2

VOLTS

F1

NEUT-GND

F2

Hot to Ground

Enter menu commands:

HOT-GND

F3

MANUAL

F2

VOLTS

F1

Leakage Current
All leakage current measurements are taken through a 1–k

Ω AAMI load RMS

or DC. Current is measured in a range of 2000

µA or 200 µA.

Note

These tests are described using Kelvin cables for connections.
However, single-wire test cables can be used. Connect single-
wire test cables to the external input posts only.

Note

For all leakage current measurements, except ECG lead
isolation,

DC

toggles the measurement between RMS and DC.