Communications settings, Setting baud rates – Fluke Biomedical medTester 5000C User Manual

Page 176

Advertising
background image

medTester 5000C
Operators Manual

11-12

Posttest Prompts

Posttest prompts that have been switched on appear upon completion of the
checklist. After responding to any posttest prompts, the medTester asks you
to store or print the test record.

Checklist Records

The number of checklist records that the medTester 5000C creates is
dependent upon the structure of the checklist. One record is stored for the
checklist itself. Individual records for autosequences executed as part of the
checklist are generated separately. The checklist record itself is generated last.
After you execute checklists that you have loaded into the medTester, you can
upload the data recorded in memory to your EMS or CMMS. In addition, you
can view or print records using the memory functions.

If elapsed test time is turned on, time is recorded for the entire checklist. If any
autosequences are executed as part of the checklist, autosequence elapsed
times are displayed as:

XXX

.

View or Print a Checklist
You can view or print a checklist (not the test record) from the checklist
menu using the same procedure described above for finding and running a
checklist. Once you select a checklist medTester displays:

Press either

VIEW

or

PRINT

.

To view the selected checklist, choose

VIEW

from the

CHECK

menu. The

example below displays how one checkitem appears when you view it:

What you see is the number of the first

ITEM

in the checklist with its sequence

number. Also the

TYPE

of checkitem appears.

TYPE

means the control type

placed on the checkitem within the EMS or CMMS when you write the
checklist.

CHECKLIST: 000001

PRINT

F5

VIEW

F4

RUN

F3

ITEM: X

TYPE: PASS/FAIL/DATA

LIGHTS & INDICATORS…CHECK OPERATION