Fluke Biomedical medTester 5000C User Manual

Page 227

Advertising
background image

Biomedical Test System

ESU Module

14

14-17

Expected Power: Enter the power expected to be generated
by the ESU, from 0 to 500 W.

Power high limit (00% = off)

Power low limit (00% = off)

RF Leakage Test Type

Electrode: Select either Active or Dispersive electrodes to
test.

Load Status: Select Open or Closed. For RF303

RS

, use only

Open Load Status.

Load (if Load Status Closed): Select an internal load (same as
for Generator Output).

Mode: Same as for Generator Output

Output: Same as for Generator Output

Power Setting: Same as for Generator Output

Contact Quality Monitor (CQM) Test Type

Nothing to select.

Return Fault Test Type (not applicable to RF303

RS

)

Test Side: Select either Cut- or Coag-side ESU mode

Power Setting: Same as for Generator Output

After each step selection, you can press

MORE

F4

to select more steps, or

END

F5

to end your selections.

After customizing your autosequence, the medTester prompts you to store
the autosequence: