Rf leakage test type, Contact quality monitor (cqm) test type, Return fault test type (not applicable to rf303rs) – Fluke Biomedical medTester 5000C User Manual

Page 223

Advertising
background image

Biomedical Test System

ESU Module

14

14-13

Autosequence Number: A63

Autosequence Name: VLF4BF

Step and Test

Type

Load

ESU Mode

Output

Power

Setting

Expected Power

or

Electrode

Limits

1

Generator Output

300 Ohms

Monocoag

Spray

50

50 W

± 20%

2

Generator Output

300 Ohms

Monocoag

Spray

120

120 W

± 20%

3

Generator Output

300 Ohms

Monocoag

Soft

60

60 W

± 20%

4

Generator Output

300 Ohms

Monocoag

Low V

99

99 W

± 20%

5

Generator Output

300 Ohms

Monopolar

Pure Cut

50

50 W

± 20%

6

Generator Output

300 Ohms

Monopolar

Pure Cut

300

300 W

± 20%

7

Generator Output

300 Ohms

Monopolar

Blend 1

250

250 W

± 20%

8

Generator Output

300 Ohms

Monopolar

Blend 2

50

50 W

± 20%

9

Generator Output

300 Ohms

Monopolar

Blend 2

200

200 W

± 20%

10

Generator Output

300 Ohms

Monopolar

Blend 3

200

200 W

± 20%

11

Generator Output

100 Ohms

Bipolar

Standard

70

70 W

± 20%

12

Generator Output

100 Ohms

Bipolar

Precise

70

70 W

± 20%

13

Auxiliary Return

Fault Test

N/A Cut

Side N/A 300

N/A

N/A

14

Auxiliary Return

Fault Test

N/A Coag

Side N/A

300

N/A

N/A

15

Auxiliary CQM

Test

N/A N/A N/A N/A N/A

N/A