Biomedical Test System

IV Pump Module

10

10-3

Infutest 2000 Series D

The Infutest can run three types tests: single rate tests, dual rate tests, and
PCA (patient controlled analgesia) tests. Even though the Infutest 2000 series
D can run up to 4 channels simultaneously, the medTester 5000C is only able
to control one channel. The medTester 5000C always controls Infutest 2000
Series D channel A to run a test.

IPT-1

With the IPT-1, you can perform 2 types of flow tests: steady flow and high
accuracy.

IPT-MC

The IPT-MC only has 1 type of flow test. Even though the IPT-MC can run
up to 4 channels simultaneously, the medTester is only able to control one
channel. The medTester 5000C always controls IPT-MC channel 1 to run a
test.


Terminating A Flow Rate Test

You can terminate a flow rate test automatically or manually. Automatic
termination occurs after a preset time has elapsed or after a preset volume has
been measured. The advantage of an automatic test termination is that you do
not need to constantly monitor the progress of each flow rate measurement.
The medTester beeps when the test is complete. In addition to having a test
automatically terminated, you can manually end a test at any time during the
flow rate measurement.