Biomedical Test System

IV Pump Module

10

10-15

• Wait for the IDA 4 Plus to display

PRIME

on channel 1. Using a priming

syringe, prime IV line into IDA 4 Plus until the IDA 4 Plus displays

• After pressing

F5

(

START

), the medTester 5000C displays

START FLOW

FOR IDA 4P AUTO-START.

• Next the IDA 4 Plus displays

AUTO

and waits for the flow to start. After

starting the flow on the IV Pump, the IDA 4 Plus should display

FLOW

followed shortly by flow data.

• If the IDA 4 Plus displays

ERROR

during a test, it is likely that a bubble

was detected in the line, and the test is invalid. End the test, and repeat it.

Single Rate Test

• For a single rate test, the medTester displays the elapsed time (ET=), the

remaining time (RT=) for a timed test, the remaining volume (RV=) for a
volume ended test, the average flow rate (A=), the instantaneous flow rate
(I=), and the cumulative volume (V=).

• After 5 minutes, the medTester checks each average flow rate value with

the programmed limits. If the average flow rate is outside the limits, it is
displayed with a ‘#’.

• You can wait for the test to end automatically, if it is a timed or volume

ended test, or you can end the test by pressing F5 END.

• At the end of the test, the medTester again checks the average flow rate

with the programmed limits. If the average flow rate is outside the limits,
it is displayed with a ‘#’ with the opportunity offered to continue or
repeat the flow test.

• The medTester displays the final flow rate test results. Press F5 NEXT.

Then you can repeat the test or continue on to the pressure test, if
selected.

Dual Rate Test

• The dual rate test measures the delivery of two sequential fluid volumes

pumped at different flow rates by the IV pump. This fluid delivery
technique is typically referred to as “piggybacking” in the clinical
environment. NOTE: the initial higher fluid delivery (“Secondary”) must
be five times (5:1) the following lower (“Primary”) flow rate and must
minimally consist of 1.0 ml measured IV fluid