To set up and connect a label printer, Selecting a label printer – Fluke Biomedical medTester 5000C User Manual

Page 80

Advertising
background image

medTester 5000C
Operators Manual

5-10

ECG Performance Waveforms

If you have the Waveform/Extended Testing module installed on your
medTester 5000C, and you have performed a lead test, this message appears:

This is a prompt to output an autosequence of performance waves to a
patient monitor from the patient leads attached to the medTester 5000C.
Waveforms are output in the order described in the table below for various
periods of time.

Waveform Name

Test Duration

(In Seconds)

2 Hz Square

Continuous

4-second Pulse

8

10 Hz Sine

4

40 Hz Sine

4

60 Hz Sine

60

100 Hz Sine

4

1 kHz Square

4

2 Hz Triangle

Continuous

CMRR 60 Hz

Continuous

ECG 60 BPM

Continuous

ECG 30 BPM

30

ECG 60 BPM

30

ECG 120 BPM

30

ECG 240 BPM

30

Note

Tests with a predefined duration automatically end. However,
you can choose to terminate them at any time by pressing

NEXT

.

Those marked continuous can be terminated by pressing

NEXT

.

OUTPUT PERFORMANCE WAVES?