medTester 5000C
Operators Manual

16-16

Begin by accessing the custom, autosequence, pacer menu as described earlier.
This time, select:

MAKE

F1

Next select the name of the autosequence you want to customize or create.
Follow the following steps, providing the desired information:

1.

Name—This is the name you want to give to the autosequence,
with a maximum of six characters.

2.

Menu Steps:

Load: Select internal 50 ohms or a TQA module

Pulse tests, 20 steps:

Current: 10 to 250 mA

Pulse Rate: 30 to 200 PPM

Load, if using a multiple test load TQA module

Current +/- limit: 0 to 99% (0% is off)

Pulse Rate +/- limit: 0 to 99 PPM (0 is off)

Demand Test:

Pacer Rate: 30 to 200 PPM

Underdrive ECG Rate: 30, 60, 80, 120, or 160 BPM

Overdrive ECG Rate: 60, 80, 120, 160, or 200 BPM

Async Test:

Pacer Rate: 30 to 200 PPM

Overdrive ECG Rate: 60, 80, 120, 160, or 200 BPM

Refractory Period Test

After customizing your autosequence, the medTester prompts you to store
the autosequence:

STORE

F5

Print Your Autosequence

From the pacer custom autosequences menu described above, printing any
autosequence is accomplished through this menu command:

PRINT