Biomedical Test System

Non-Invasive Blood Pressure (NIBP) Module

17

17-13

Begin by accessing the custom, autosequence, NIBP menu as described
earlier. This time, select:

MAKE

F1

Next select the name of the autosequence you want to customize or create.
Follow the following steps, providing the desired information:

1.

Name—This is the name you want to give to the autosequence,
with a maximum of six characters.

2.

Menu Steps:

NIBP Tests, 10 steps:
NIBP

selection:

60/ 30( 40) ADULT

80/ 50( 62) ADULT

100/ 65( 75) ADULT

120/ 80( 90) ADULT

150/100(115)

ADULT

200/150(165)

ADULT

255/195(215)

NEONATE

60/ 30( 40) NEONATE

80/ 50( 62) NEONATE

100/ 65( 75) NEONATE

120/ 80( 90) NEONATE

150/100(115)

NEONATE

120/ 80( 90) PAC

120/ 80( 90) PVC

120/ 80( 90) AFIB

120/ 80( 90) MBC

120/ 80( 90) ASC

Heart Rate: (except for arrhythmias)

30 to 240 BPM (80 nominal)

Envelope Gain: (except for arrhythmias)

0 to 200 % (100 nominal)

Envelope Shift: (except for arrhythmias)

Zero

Positive: 0 to 100

Negative: 0 to 100

NIBP +/- LIMIT: 0 t0 99 % (0 off, 20 nominal)

HEART RATE +/- LIMIT: 0 to 99 BPM

(0 off, 4 nominal)

PRESSURE TEST

LEAK TEST

POP OFF TEST