Generator output test type – Fluke Biomedical medTester 5000C User Manual
Page 222
medTester 5000C
Operators Manual
14-12
Autosequence Number: A62
Autosequence Name: VLF40
Step and Test
Type
Load
ESU Mode
Output
Power
Setting
Expected Power
or
Electrode
Limits
1
Generator Output
300 Ohms
Monocoag
Spray
150
150 W
± 15%
2
Generator Output
300 Ohms
Monopolar
Pure Cut
300
300 W
± 15%
3
Generator Output
300 Ohms
Monopolar
Blend 1
250
250 W
± 15%
4
Generator Output
300 Ohms
Monopolar
Blend 2
200
200 W
± 15%
5
Generator Output
300 Ohms
Monopolar
Dessicate
200
200 W
± 15%
6
Generator Output
300 Ohms
Monopolar
Fulgurate
150
150 W
± 15%
7
RF Leakage
Open Monocoag
Dessicate 200 Active
Electrode N/A
8
RF Leakage
Open Monocoag
Dessicate 200
Dispersive
Electrode
N/A
9
RF Leakage
Open
Monopolar
Pure Cut
300
Active Electrode
N/A
10
RF Leakage
Open Monopolar Pure
Cut 300
Dispersive
Electrode
N/A
11
Generator Output
100 Ohms
Bipolar
Standard
50
50 W
± 15%
12
Generator Output
100 Ohms
Bipolar
Precise
50
50 W
± 15%