Generator output test type – Fluke Biomedical medTester 5000C User Manual

Page 222

Advertising
background image

medTester 5000C
Operators Manual

14-12

Autosequence Number: A62

Autosequence Name: VLF40

Step and Test

Type

Load

ESU Mode

Output

Power

Setting

Expected Power

or

Electrode

Limits

1

Generator Output

300 Ohms

Monocoag

Spray

150

150 W

± 15%

2

Generator Output

300 Ohms

Monopolar

Pure Cut

300

300 W

± 15%

3

Generator Output

300 Ohms

Monopolar

Blend 1

250

250 W

± 15%

4

Generator Output

300 Ohms

Monopolar

Blend 2

200

200 W

± 15%

5

Generator Output

300 Ohms

Monopolar

Dessicate

200

200 W

± 15%

6

Generator Output

300 Ohms

Monopolar

Fulgurate

150

150 W

± 15%

7

RF Leakage

Open Monocoag

Dessicate 200 Active

Electrode N/A

8

RF Leakage

Open Monocoag

Dessicate 200

Dispersive

Electrode

N/A

9

RF Leakage

Open

Monopolar

Pure Cut

300

Active Electrode

N/A

10

RF Leakage

Open Monopolar Pure

Cut 300

Dispersive

Electrode

N/A

11

Generator Output

100 Ohms

Bipolar

Standard

50

50 W

± 15%

12

Generator Output

100 Ohms

Bipolar

Precise

50

50 W

± 15%