medTester 5000C
Operators Manual

16-14

Pacer Autosequence Number: A24

Autosequence Name: ZOLLPM

Load: Adapter Module TQA-2, External 1000 Ohms

Step Number

Current

(in milliamperes)

Rate

(in PPM, pulses per minute)

1 80

180

2 80

160

3 80

140

4 80

120

5 80

100

6 80

80

7 80

60

8 80

40

9 140

80

10 100

80

11 60

80

12 20

80

Pulse Tests:

Current Limit: ± 20%

Rate Limit: ± 5 PPM

Demand Test: 80 PPM

Underdrive: 30 BPM

Ovedrive: 160 BPM

Async Test: No
Refractory Period Test: No

Posttest Prompts
When all tests are completed, the medTester 5000C prompts you for:

1.

COMMENTS

2.

NEXT TEST DUE DATE

3.

USER TIME

EMS/CMMS system users:

User time is required

4.

PRINT TEST TAG

5.

STORE or PRINT TEST RECORD