Medcheck module, Tap your medtester™s potential – Fluke Biomedical medTester 5000C User Manual

Page 162

Advertising
background image

medTester 5000C
Operators Manual

10-24

2.

SECONDARY RATE HIGH LIMIT: Enter the high limit for the
flow rate. 0000.0 turns off high limit checking.

3.

SECONDARY RATE LOW LIMIT: Enter the low limit for the
flow rate. 0000.0 turns off low limit checking.

4.

PRIMARY FLOW RATE: Enter the nominal flow rate, 0000.1 to
0999.0 ml/hr.

5.

PRIMARY RATE HIGH LIMIT: Enter the high limit for the
flow rate. 0000.0 turns off high limit checking.

6.

PRIMARY RATE LOW LIMIT: Enter the low limit for the flow
rate. 0000.0 turns off low limit checking.

7.

TEST END TYPE: Select

8.

MANUAL to end the test by pressing a key.

9.

TIMED to end the test after a specific time:

10.

END TIME: Enter the elapsed time that will end the
test, 00:01 to 99:59 (hours:minutes).

11.

VOLUME to end the test after a specific volume:

12.

END VOLUME: Enter the volume that will end the
test, 0000.1 to 9999.0 ml.

For a PCA Test:

• BOLUS VOLUME: Enter the bolus volume, 00.50 to 99.99 ml.

• BOLUS VOLUME HIGH LIMIT: Enter the high limit for the

bolus volume. 00.00 turns off high limit checking.

• BOLUS VOLUME LOW LIMIT: Enter the low limit for the

bolus volume. 00.00 turns off low limit checking.

• LOCKOUT TIME: Enter the lockout time between boluses, 00:01

to 99:99 (minutes:seconds).
• END TIME: Enter the elapsed time that will end the test, 00:01

to 99:99 (hours:minutes).

IPT-1

For the IPT-1, there are 2 types of tests available, steady flow and high
accuracy:

1.

FLOW RATE: Enter the nominal flow rate, 0000.1 to 0999.0
ml/hr.

2.

FLOW TEST TYPE: Enter STEADY for steady flow or HI ACC
for high accuracy.

3.

RATE HIGH LIMIT: Enter the high limit for the flow rate.
0000.0 turns off high limit checking.