Ipt-mc, Print your autosequence – Fluke Biomedical medTester 5000C User Manual

Page 159

Advertising
background image

Biomedical Test System

IV Pump Module

10

10-21

IPT-MC AUTOSEQUENCE TESTS

Nominal

Flow Rate

(ml per

hour)

High
Limit

(ml per

hour)

Low

Limit

(ml per

hour)

End Type

(manual,

volume,

or timed)

Ending Quantity

If Applicable

(ml for volume or

hours:minutes

for timed)

Autosequence Name (Test Type): PUMP 1

Keyboard Shortcut Command: A50

60.0 66.0 54.0

Volume

5.0

120.0 132.0 108.0

Volume

10.0

240 264.0

216.0

Volume 20.0

Autosequence Name (Test Type): PUMP 2

Keyboard Shortcut Command: A51

30.0 35.0 25.0

Volume

3.0

60.0 66.0 54.0

Volume

5.0

120.0 132.0 108.0

Volume

10.0

120.0 132.0 108.0

Volume

10.0

Autosequence Name (Test Type): LOW

Keyboard Shortcut Command: A52

10.0 11.0 9.0 Timed 00:05

Autosequence Name (Test Type): HIGH

Keyboard Shortcut Command: A53

480.0 504.0 456.0 Timed

00:05

Autosequence Name (Test Type): EVAL

Keyboard Shortcut Command: A54

5.0 5.9 4.1

Volume 1.5

10.0 11.0 9.0

Volume

2.5

20.0 22.0 18.0

Volume

2.5

30.0 33.0 27.0

Volume

2.5

60.0 66.0 54.0

Volume

5.0

120.0 132.0 108.0

Volume

10.0

240.0 264.0 216.0

Volume

30.0

480.0 528.0 432.0

Volume

60.0

240.0 264.0 216.0

Volume

30.0

480.0 528.0 432.0

Volume

60.0