Customizing pacer autosequences, Make your own autosequence – Fluke Biomedical medTester 5000C User Manual
Page 267
Advertising
Biomedical Test System
Transcutaneous Pacemaker (Pacer) Module
16
16-11
Pacer Autosequence Number: A21
Autosequence Name: LP9PM
Load: Adapter Module TQA-16, External 700 Ohms
Step Number
Current
(in milliamperes)
Rate
(in PPM, pulses per minute)
1 55
170
2 55
160
3 55
150
4 55
140
5 55
130
6 55
120
7 55
110
8 55
100
9 55
90
10 55
80
11 55
70
12 55
60
13 55
50
14 55
40
15 200
60
16 150
60
17 130
60
18 100
60
Pulse Tests:
• Current Limit: ± 20%
• Rate Limit: ± 5 PPM
Demand Test: 60 PPM
• Underdrive: 30 BPM
• Ovedrive: 160 BPM
Async Test: 60 PPM
• Overdrive: 160 BPM
Refractory Period Test: No