Customizing pacer autosequences, Make your own autosequence – Fluke Biomedical medTester 5000C User Manual

Page 267

Advertising
background image

Biomedical Test System

Transcutaneous Pacemaker (Pacer) Module

16

16-11

Pacer Autosequence Number: A21

Autosequence Name: LP9PM

Load: Adapter Module TQA-16, External 700 Ohms

Step Number

Current

(in milliamperes)

Rate

(in PPM, pulses per minute)

1 55

170

2 55

160

3 55

150

4 55

140

5 55

130

6 55

120

7 55

110

8 55

100

9 55

90

10 55

80

11 55

70

12 55

60

13 55

50

14 55

40

15 200

60

16 150

60

17 130

60

18 100

60

Pulse Tests:

Current Limit: ± 20%

Rate Limit: ± 5 PPM

Demand Test: 60 PPM

Underdrive: 30 BPM

Ovedrive: 160 BPM

Async Test: 60 PPM

Overdrive: 160 BPM

Refractory Period Test: No