For the oxitest plus – Fluke Biomedical medTester 5000C User Manual

Page 241

Advertising
background image

Biomedical Test System

SPO2 Module

15

15-13

For the Oxitest PLUS:

SPO2 Autosequence Number: A70

Autosequence Name: ADULT1

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step Number

SPO2

( in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

Artifact

1 97 ±4

75 100 Off

2 97 ± 4

45 100 Off

3 97 ± 4

180 100 Off

4 80 ± 4

45 100 Off

5 90 ± 4

120 100 Off

6 80 ± 4

180 100 Off

SPO2 Autosequence Number: A71

Autosequence Name: ADULT2

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step

Number

SPO2

( in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

Artifact

1 97 ±4

75 100 Off

2 97 ± 4

45 100 Off

3 97 ± 4

180 100 Off

4 80 ± 4

45 100 Off

5 90 ± 4

120 100 Off

6 80 ± 4

180 100 Off

Starting Alarm Value for SPO2: 93%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–97% Lo–80%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test