For the index 2xl, For the oxitest plus – Fluke Biomedical medTester 5000C User Manual

Page 243

Advertising
background image

Biomedical Test System

SPO2 Module

15

15-15

For the Oxitest PLUS:

SPO2 Autosequence Number: A74

Autosequence Name: NEO2

Settling Time: 25 Beats

Heart Rate Limit:

±6 BPM (beats per minute)

Step

Number

SPO2

(in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

Artifact

1 93 ± 4

145

40

Off

2 93 ± 4

75

40

Off

3 93 ± 4

225

40

Off

4 97 ± 4

225

40

Off

5 80 ± 4

60

40

Off

6 93 ± 4

145

40

Movement

7 93 ± 4

145

40

Tremor

Starting Alarm Value for SPO2: 93%
Starting Alarm Value for Rate: 145 BPM
Starting Alarm Value for Amplitude: 40%
SPO2 Alarm Test: Hi–97% Lo–90%
Rate Alarm Test: Hi–165 BPM Lo–95 BPM
Asystole Test
Amplitude Test
Probe Test