Power cord resistance – Fluke Biomedical medTester 5000C User Manual

Page 51

Advertising
background image

Biomedical Test System

Manual Tests

4

4-5

Lead isolation, all leads.

Lead isolation, single lead

Note

For all lead-to-ground and lead-to-lead leakage measurements,
you can configure the test receptacle with open or closed
ground, normal or reverse polarity, and open or closed neutral
by toggling

GROUND, POLARITY,

or

NEUTRAL

on the display.

Leakage to Ground—All Leads

See Figure 4-5, ECG Lead Leakage to Ground: All Leads to Ground and
Individual Lead to Ground.

1.

Plug the EUT into the

TEST RECEPTACLE.

2.

Connect all ECG leads to their respective binding posts on the
medTester 5000C.

3.

Enter menu commands:

ALL/GND

F1

µA mV

F2

MANUAL

F2

ECG

F4

Displayed measurements are in

µA.

Leakage to Ground—Individual Lead

See Figure 4-5, ECG Lead Leakage to Ground: All Leads to Ground and
Individual Lead to Ground.

1.

Plug the EUT into the

TEST RECEPTACLE.

2.

Connect the ECG lead to be tested to its respective binding post
on the medTester 5000c.

Enter menu commands:

MANUAL

F2

µA

µA

F2

ECG

F4

LD/GND

F2

Select the desired lead. Lead names are: