Remote commands list – Fluke Biomedical medTester 5000C User Manual

Page 190

Advertising
background image

medTester 5000C
Operators Manual

12-12

Remote Commands List

Commands sent to the medTester from a remote computer include:

General Administrative Commands,

File Transfer Commands,

Setup Commands,

Measurement Commands, and

Wave Commands.

GENERAL ADMINISTRATIVE COMMANDS

Command

Explanation

REMOTE

Begins a remote mode session on the
medTester 5000C.

REMOTE1

Begins a remote mode session on the medTester’s
COM1 port.

REMOTE2

Begins a remote mode session on the medTester’s
COM2 port.

LOC

Ends a remote session and returns the
medTester 5000C to local mode.

IDENT

Asks the medTester for the instrument
identification name, firmware version level, and any
installed options separated by commas. Options
can include: BARCODE; RS232/PRINTER; 100
RECORDS; EXPANDED MEMORY;
WAVEFORM/EXTENDED TESTING; DATA
TRANSFER; DEFIB, IPT; IV PUMP, IDA; ESU; SP02;
EPA; NIBP; MEDCHECK.

SERN

Returns the medTester serial number in five (5)
digits.

REV

Returns the firmware version number.

SETCLOCK=

SETCLOCK sets the real time clock. Time is in a 24-
hour format. Follow these conventions:

Formatted as: Month, day, year, hour, minute,
second. Separate each entry with a comma. Do not
include spaces.

For example, using the listed conventions, here is
how you set January 1, 1997, 2:11

P

.

M

. and 14

seconds

1,1,97,14,11,14


month: an integer 1 through 12
day: an integer 1 through 31
year: an integer 0 through 99
hour: an integer 0 through 23
minute: an integer 0 through 59
second: an integer 0 through 59.