End-of-test prompts, Toolbox monitors, Temperature monitor – Fluke Biomedical medTester 5000C User Manual

Page 85: Optional monitoring autosequences

Advertising
background image

Biomedical Test System

Autosequences

5

5-15

1.

From the menu, press

F1

(

CONFIG

) as before to configure your

test tag (test tag lines on/off), or press

F2

(

PRINTER

) to select a

label printer.

2.

From the (

PRINTER

) menu press

F1

(

TLP

) to select Fluke

Biomedical TLP printers (TLP-1 or TLP-2), or

F2

(

BRADY

) to

select the Brady printer. The selection is displayed.

3.

Press

F5

(

STORE

) to store the selection in nonvolatile

EEPROM memory.

Note

When using the Brady printer, it is not necessary to set the
number of lines on the label with the

UTIL / PRTPAGE / TAGS

menu path. However, you still need to select the fields you want
printed on the label through the

CUSTOM

/

TAG

/

CONFIG

menu

path.

To configure a medTester serial port for test tag printing, see
Local Output for Records and Tags in Chapter 12, Remote
Operation
. To configure one of the Wedge adapter COM2 ports
for test tag printing, see Setting Output Ports in Chapter 13, The
Wedge Adapter

Optional Monitoring Autosequences

Two optional autosequences are available in the Waveform/Extended Testing
Module. Both are designed as monitors.

Line monitor voltage test—periodically measures line voltage, and

Environmental test—which measures ground potential and
resistance between a common ground point and up to 99 other
ground points in a room.

MONITORING AUTOSEQUENCES

Autosequence

Name

Function

Menu Command

From MENU 2

Remote

Command

LINEMON Line

voltage

monitor.

LINEMON

F4

AUTO

F1

A16

ENVMON Ground

potential

and resistance.

ENVMON

F5

AUTO

F1

A17