Infutest, For a single rate test, For a dual rate test – Fluke Biomedical medTester 5000C User Manual

Page 157

Advertising
background image

Biomedical Test System

IV Pump Module

10

10-19

INFUTEST 2000 Series D AUTOSEQUENCE TESTS

Nominal

Flow Rate

(ml per

hour)

Test Type

High
Limit

(ml per

hour)

Low

Limit

(ml per

hour)

End Type

(manual,

volume,

or timed)

Ending Quantity

If Applicable

(ml for volume or

hours:minutes

for timed)

Autosequence Name (Test Type): PUMP 1

Keyboard Shortcut Command: A50

60.0 Single

Rate

66.0 54.0 Volume

5.0

120.0 Single

Rate

132.0 108.0 Volume

10.0

240 Single

Rate

264.0 216.0 Volume

20.0

Autosequence Name (Test Type): PUMP 2

Keyboard Shortcut Command: A51

30.0 Single

Rate

35.0 25.0 Volume

3.0

60.0 Single

Rate

66.0 54.0 Volume

5.0

120.0 Single

Rate

132.0 108.0 Volume

10.0

240.0 Single

Test

264.0 216.0 Volume

20.0

Autosequence Name (Test Type): DUAL

Keyboard Shortcut Command: A52

Secondary

120.0

Primary

20.0

Dual Rate

132.0

25.0

108.0

15.0

Manual

Autosequence Name (Test Type): PCA

Keyboard Shortcut Command: A53

Bolus Volume

1.0ml

Lockout Time

30.00(mm:ss)

PCA 1.30ml .70ml Timed

03.00

Autosequence Name (Test Type): EVAL

Keyboard Shortcut Command: A54

5.0 Single

Rate

5.9 4.1 Volume

1.5

10.0 Single

Rate

11.0 9.0 Volume

2.5

20.0 Single

Rate

22.0 18.0 Volume

2.5

30.0 Single

Rate

33.0 27.0 Volume

2.5

60.0 Single

Rate

66.0 54.0 Volume

5.0

120.0 Single

Rate

132.0 108.0 Volume

10.0

240.0 Single

Rate

264.0 216.0 Volume

30.0

480.0 Single

Rate

528.0 432.0 Volume

60.0

Advertising