Iv pump module, Iv pump autosequences, Configuring for your iv pump analyzer – Fluke Biomedical medTester 5000C User Manual

Page 135

Advertising
background image

Biomedical Test System

Defibrillator Module

9

9-11

The table below supplies specific information about the tests for each
autosequence.

Autosequence

Name

Energy

Step High

Limit

(%)

Energy

Step Low

Limit

(%)

Max

Energy

High

Limit

(Joules)

Max

Energy

Low Limit

(Joules)

Paddle

Leakage

and/or

Cardioversion

Energy Steps

(Joules)

LP4

10

10

480

418

Cardioversion

10, 25, 50, 100,

200, 300, 400, 450

LP5

10

10

396

324

Cardioversion

20, 50, 100, 200,

300, 360

LP6

10

10

428

372

Both

5, 10, 20, 30, 50,

100, 200, 300, 400

LP6S

10

10

385

335

Both

5, 10, 29, 30, 50,

100, 150, 200, 300,

360

LP7

10

10

385

335

Both

5, 10, 20, 30, 50,

100, 200, 300, 360

LP8

10

10

385

335

Both

2, 5, 9, 10, 20, 30,
50, 100, 150, 200,

300, 360

M/D3

10

10

440

360

Both

5, 10, 20, 40, 60,

100, 150, 200, 250,

300, 360

78660A

10

10

400

330

Both

5, 10, 20, 40, 60,

100, 150, 200, 250,

300, 360

NK7000

10

10

396

324

Both

3, 5, 10, 20, 30, 50,

70, 100, 150, 200,

300, 360

AOLOWN

15

15

414

306

Both

5, 10, 20, 30, 50,

100, 150, 200, 300,

360

M/D4

20

20

384

256

Both

1, 8, 40, 80, 160,

240, 320

M/D2J

10

10

506

404

Both

5, 10, 20, 40, 75,

100, 200, 300, 400,

460

78670A

15

15

414

306

Both

5, 10, 20, 30, 50,

70, 100, 150, 200,

300, 360

43100A

15

15

414

306

Both

2, 3, 5, 7, 10, 30,
50, 70, 100, 150,

200, 300, 360

ZOLL

10

10

396

324

Both

2, 3, 5, 7, 10, 30,

50, 100, 150, 200,

300, 360

LP10

10

10

396

324

Cardioversion

5, 10, 20, 50, 100,

200, 300, 360

3 blanks

(user-defined)