Fluke Biomedical medTester 5000C User Manual

Page 94

Advertising
background image

medTester 5000C
Operators Manual

6-2

5.

After pressing this function key, you can begin scrolling through
the steps in the sequence by using your up and down arrow keys.
Continue viewing our example sequence below.

What You See in an Autosequence

As you scroll through autosequence A1, you see the following:

The steps we are viewing are the leakage current tests. This list is not
complete; view the entire autosequence for the remaining ones. There are 62
different leakage current tests in the medTester 5000C. Note that
autosequence A1 (

10LD

X) does not contain all 62 tests. You can see all tests

listed in the table Leakage Current Tests found later in this chapter.

In addition to viewing the leakage current tests in this autosequence, you see a
statement about line voltage tests and the four test limits programmed into
this autosequence. Since autosequence configurations may differ from one
autosequence to another, you could view statements for power-off
measurements and single-step.

1 CASE EXT, NORM POL, CLSD GND, CLSD NEU
2 CASE EXT, NORM POL, OPEN GND, CLSD NEU
13 ALL -GND, NORM POL, CLSD GND, CLSD NEU
14 RL -GND, NORM POL, CLSD GND, CLSD NEU