Comments, Next test due date, User time – Fluke Biomedical medTester 5000C User Manual

Page 285: User time is required, Print test tag, Store or print test record

Advertising
background image

Biomedical Test System

Non-Invasive Blood Pressure (NIBP) Module

17

17-11

NIBP Autosequence Number: A83

Autosequence Name: NEONT1

Step Number

Blood Pressure

Heart Rate in BPM
(beats per minute)

Envelope Gain

Envelope Shift

1

80/50 (62) NEONT

120

100%

+0

2

60/30 (40) NEONT

80

100%

+0

3

100/65 (75) NEONT

180

100%

+0

NIBP Test Limits:

NIBP: ± 20%

Heart Rate: ± 6 BPM

Pressure Test: No
Leak Test: No
Pop Off Test: No

NIBP Autosequence Number: A84

Autosequence Name: NEONT2

Step Number

Blood Pressure

Heart Rate in BPM
(beats per minute)

Envelope Gain

Envelope Shift

1

80/50 (62) NEONT

120

100%

+0

2

60/30 (40) NEONT

80

100%

+0

3

100/65 (75) NEONT

180

100%

+0

NIBP Test Limits:

NIBP: ± 20%

Heart Rate: ± 6 BPM

Pressure Test: Yes
Leak Test: Yes
Pop Off Test: Yes

Posttest Prompts
When all tests are completed, the medTester 5000C prompts you for:

1.

COMMENTS

2.

NEXT TEST DUE DATE

3.

USER TIME

EMS/CMMS system users:

User time is required.

4.

PRINT TEST TAG

5.

STORE or PRINT TEST RECORD