medTester 5000C
Operators Manual
16-10
Pacer Autosequence Number: A20
Autosequence Name: LP12P
Load: Internal 50 Ohms
Step Number
Current
(in milliamperes)
Rate
(in PPM, pulses per minute)
1 10
40
2 10
60
3 50
60
4 100
60
5 150
60
6 200
60
7 200
40
8 200
170
Pulse Tests:
• Current Limit: ± 20%
• Rate Limit: ± 5 PPM
Demand Test: 80 PPM
• Underdrive: 30 BPM
• Ovedrive: 160 BPM
Async Test: No
Refractory Period Test: No