Print your autosequence – Fluke Biomedical medTester 5000C User Manual

Page 268

Advertising
background image

medTester 5000C
Operators Manual

16-12

Pacer Autosequence Number: A22

Autosequence Name: HPCMPM

Load: Adapter Module TQA-12, External 400 Ohms

Step Number

Current

(in milliamperes)

Rate

(in PPM, pulses per minute)

1 30

40

2 30

50

3 30

60

4 30

70

5 30

80

6 30

90

7 30

100

8 30

110

9 30

120

10 30

130

11 30

140

12 30

150

13 30

160

14 30

170

15 30

180

16 60

80

17 100

80

18 140

80

19 180

80

20 200

80

Pulse Tests:

Current Limit: ± 20%

Rate Limit: ± 5 PPM

Demand Test: 80 PPM

Underdrive: 30 BPM

Ovedrive: 160 BPM

Async Test: 80 PPM

Overdrive: 160 BPM

Refractory Period Test: No