medTester 5000C
Operators Manual

12-8

UNIT of MEASUREMENT ABBREVIATIONS

Abbreviation

Unit of Measurement

V RMS

Volts RMS

A RMS

Amperes RMS

µA RMS

Microamperes RMS

µA DC

Microamperes DC

OHMS Ohms
RPM

Revolutions per minute

DEG F

Degrees Fahrenheit

DEG C

Degrees Centigrade

% Percent

Error Responses

If you issue an incorrect or nonexistent remote command, the medTester
responds with an error statement in this format:

ERR=

xx. Here is a list of

possible error responses and explanations.

ERROR RESPONSES

Error Statement

Explanation

ERR=01,UNKNOWN COMMAND

The received command was not
recognized.

ERR=02,ILLEGAL COMMAND

The received command was not
allowed for the current instrument
status.

ERR=03,ILLEGAL PARAMETER

A parameter sent with a command
was not the right type, was out of
range, or was not understood, or
not the right number of parameters
were sent.

ERR=04,MODULE NOT INSTALLED

Can not execute the command
because the required instrument
module is not installed.

ERR=20,CURRENT SOURCE NOT
CONNECTED

For a resistance measurement
command, the current source
connection has not been made.

ERR=21,MEMORY FULL

For a file down transfer command,
the medTester memory is full and
can not accept anymore data to
fulfill the transfer.

ERR=22,GFCI

A ground fault circuit interruption
was triggered caused by a fault in
the EUT plugged into the test
receptacle. This error is not a
response to a command. It
appears whenever a ground fault
occurs.

ERR=23,WRONG FILE

For a file transfer command where