Biomedical Test System

ESU Module

14

14-11

Autosequence Number: A61

Autosequence Name: VLF30

Step and Test

Type

Load

ESU Mode

Output

Power

Setting

Expected Power

or

Electrode

Limits

1

Generator Output

300 Ohms

Monocoag

Dessicate

200

200 W

± 15%

2

Generator Output

300 Ohms

Monocoag

Fulgurate

150

150 W

± 15%

3

Generator Output

300 Ohms

Monopolar

Pure Cut

300

300 W

± 15%

4

Generator Output

300 Ohms

Monopolar

Blend 1

250

250 W

± 15%

5

RF Leakage

Open Monocoag

Dessicate 200 Active

Electrode N/A

6

RF Leakage

Open Monocoag

Dessicate 200

Dispersive

Electrode

N/A

7

RF Leakage

Open

Monopolar

Pure Cut

300

Active Electrode

N/A

8

RF Leakage

Open Monopolar Pure

Cut 300

Dispersive

Electrode

N/A

9

Generator Output

100 Ohms

Bipolar

Standard

50

50 W

± 15%

10

Generator Output

100 Ohms

Bipolar

Precise

50

50 W

± 15%