Customizing esu autosequences, Make your own autosequence – Fluke Biomedical medTester 5000C User Manual
Page 221
Biomedical Test System
ESU Module
14
14-11
Autosequence Number: A61
Autosequence Name: VLF30
Step and Test
Type
Load
ESU Mode
Output
Power
Setting
Expected Power
or
Electrode
Limits
1
Generator Output
300 Ohms
Monocoag
Dessicate
200
200 W
± 15%
2
Generator Output
300 Ohms
Monocoag
Fulgurate
150
150 W
± 15%
3
Generator Output
300 Ohms
Monopolar
Pure Cut
300
300 W
± 15%
4
Generator Output
300 Ohms
Monopolar
Blend 1
250
250 W
± 15%
5
RF Leakage
Open Monocoag
Dessicate 200 Active
Electrode N/A
6
RF Leakage
Open Monocoag
Dessicate 200
Dispersive
Electrode
N/A
7
RF Leakage
Open
Monopolar
Pure Cut
300
Active Electrode
N/A
8
RF Leakage
Open Monopolar Pure
Cut 300
Dispersive
Electrode
N/A
9
Generator Output
100 Ohms
Bipolar
Standard
50
50 W
± 15%
10
Generator Output
100 Ohms
Bipolar
Precise
50
50 W
± 15%