Serial port expansion, Keyboard interface, Port configuration – Fluke Biomedical medTester 5000C User Manual

Page 194

Advertising
background image

medTester 5000C
Operators Manual

12-16

WAVE COMMANDS

Note: The following commands turn on the specified wave. The wave continues to run until it
is changed by another wave command, or it will be turned off and disconnected by any non-
wave command or the EXIT command.

All wave commands return OK.

Command

Explanation

W000 Zero

output.

W001

2 Hz square wave.

W002

4 second pulse.

W003

10 Hz sine wave.

W004

40 Hz sine wave.

W005

60 Hz sine wave.

W006

100 Hz sine wave.

W007

1 kHz square wave.

W008

2 Hz triangle wave.

W009

60 Hz common mode rejection test.

W010

Normal Sinus ECG at 30 BPM.

W011

Normal Sinus ECG at 60 BPM.

W012

Normal Sinus ECG at 120 BPM.

W013

Normal Sinus ECG at 240 BPM.

W014

0.5 Hz sine wave.

W015

Atrial fibrillation 1, course.

W016

2nd degree AV block, type 1, Wenckebach.

W017

2nd degree AV block, type 2.

W018

Missed beat at 80 BPM.

W019

Missed beat at 120 BPM.

W020

PVC type 1, left.

W021

PVC type 2, right.

W022 Multifocal

PVCs.

W023

PVC 1, R on T.

W024

A pair or couplet of PVCs.

W025

Run of 5 PVCs.

W026

Run of 11 PVCs.

W027

Right bundle branch block.

W028

Ventricular tachycardia 180 BPM.

W029 Ventricular

fibrillation.

W030

Asystole 2, flat.