Yaskawa VS-626 MC5 User Manual

Page 175

Advertising
background image

7.3

Common Functions

- 37

Sample/hold
command

Analog input

Frequency
reference

CLOSED

100

ms

100

ms

100

ms

CLOSED

CLOSED

Fig

7.21

Analog Frequency Reference Sample/Hold

D

The analog frequency reference sample/hold function is valid only for terminals 13, 14, and 16 or for
the analog inputs from the AI-14U or AI-14B.

D

An OPE03 fault will occur if two or more of the following signals turn ON at the same time: accelera-
tion/deceleration ramp hold command (0A), up/down commands (10 or 11), trim control increase/de-
crease commands (1C or 1D), and the analog frequency reference sample/hold command.

External Faults (Settings: 20 to 2F)

D

With this setting, the multi-function input can be used to stop the Inverter or output an alarm when a
malfunction or fault occurs in a peripheral device.

D

There are 16 external fault inputs available with all 16 combinations of the following variables. Select
the setting with the desired combination.

x

Input level:

Normally open or normally closed

x

Detection method:

Always or During operation only

x

Operation selection:

Deceleration to stop, Coast to stop, Emergency stop, or Continue opera
tion

Table

7.6

External Fault Settings

Input level

Detection method

Operation selection

Setting

NO con-

tact

NC con-

tact

Always

During

operation

Decel-

eration to

stop

(Fault)

Coast to

stop (Fault)

Emer-
gency

stop

(Fault)

Continue

operation

(Alarm)

20

21

22

23

24

25

26

27

28

29

2A

2B

2C

2D

2E

2F

D

For the input level, select whether you want a fault to be detected when the input signal is ON (normally
open input) or OFF (normally closed input).

D

For the detection method, select whether you want faults to be detected any time that the Inverter is
ON or only during operation.

7

Advertising