Epc program setting – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 108

Advertising
background image

<1. Overview>

1-81

IM 11B08A01-01E

EPC program setting

When pressure adjustor for career gas complies with EPC specifi cation of program control,
three-staged heating program can be set.

Time

Warm up time

for stream 2

Stream 1

Start

Start

Analysis cycle

Stream 2

Analysis cycle

Pressure

3rd control

pressure

2nd control

pressure

1st control

pressure

Initial control

pressure

1st rise

2nd rise

3rd rise

Initial

control

time

1st

control

time

2nd

control

time

3rd

control

time

Figure 1.39

EPC program operations

11/03/22

15:15:45

SYS Setup (3/4)

SYS #1 Oven #1

- Carrier 1-1

Executed

- Carrier 1-2

Executed

> -

Utility 1-1

Executed

- Utility 1-2

Executed

- Utility 1-3

----

- Utility 1-4

----


Menu

Peak

Status

Get

SYS# Oven#

F1

F2

F3

F4

F5

F6

Figure 1.40

Setting example of SYS (GC-HMI EtherLCD)

2nd Edition : May 11, 2012-00

1

Advertising