● In the case where a stream specifi cation change command (for continuous measurement)

is received during the run mode in the stream specifi cation status (for continuous
measurement) [2B2]

The measurement status is changed to the requested stream specifi cation status (for

continuous measurement) when the measurement of the currently-measured stream is
completed.

In this case, the operation mode remains unchanged from the run mode.

Run

W4

W1

S1

W1

S1

S4

W4

S4

W4

S4

W4

[2B2]

Time

Stream specification 1 (Continuous)

Stream specification 4 (Continuous)

Assinning stream specification 4 (Continuous)

● In the case where a stream specifi cation change command (for continuous measurement)

is received during the run mode in the stream specifi cation status (for N times of
measurement) [3B2]

The measurement status is changed to the requested stream specifi cation status (for

continuous measurement) when the Nth round of the measurement of the currently-
measured stream is completed.

In this case, the operation mode remains unchanged from the run mode.

Stop

Run

W4

W1

S1

W1

S1

W1

S1

S4

W4

S4

W4

S4

W4

[3B2]

Time

Stream specification 1 (3 times)

Stream specification 4 (Continuous)

Assinning stream specification 4 (Continuous)

Run command

● In the case where a stream specifi cation change command (for N times of measurement)

is received during the run mode in the stream specifi cation status (for continuous
measurement) [2C2]

The measurement status is changed to the requested stream specifi cation status (for

N times of measurement) when the measurement of the currently-measured stream is
completed.

In this case, the operation mode changes from the run mode to the stop mode when the Nth

round of the specifi ed measurement is completed.

Stop

Run

W4

W1

S1

W1

S1

S4

W4

S4

[2C2]

Time

Stream specification 1 (Continuous)

Stream specification 4 (Twice)

Assinning stream specification 4 (Twice)

* Measurement status cannot be changed