8 peak setup-general – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 314

Advertising
background image

<5. EtherLCD>

5-63

IM 11B08A01-01E

5.4.8 Peak

Setup-General

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Peak Setup-General

GCM #1
>

- Conc calc req Corr area norm

- Det slope ratio 0.00
- Calib range 0.500
- Calib repeat range 2.000
- Calib coef range 4.000
- #1 Tickmark setup None
- #2 Tickmark setup None
- Peak detect err set Executed

Menu

GCM#

F1

F2

F3

F4

F5

F6

Figure 5.77

Example of the peak setup-general screen

F1(Menu):

Displays Table Menu screen

F6(GCM#): Specifi es GCM number.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #

1

6

0: Standard, 1: Correction area

fraction

Conc cal req

Det slope

ratio

0.00

9.99

Calib range

0.000

1.000

Calib repeat

range

0.000

2.000

Calib coef

range

0.000

4.000

#1 Tickmark

setup

None, Start, 10s, Start − 10s,

60s, Start − 60s, 10s to 60s,

Start −10s to 60s

#2 Tickmark

setup

None, Gate, Peak, Gate to Peak

Peak detect

err set

Not executed, Executed

2nd Edition : May 11, 2012-00