Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 157

Advertising
background image

<3. Basic Operation and Startup>

3-6

IM 11B08A01-01E

(3) The temperature information can be checked. Refer to fi gure 3.6.

F0306.ai

Temperature controller
information

Detector information

Temperature information

Pressure display information

Pressure information

(only with EPC)

Figure 3.6

Example of Oven Information Display Area

(4) When

is pressed, the set temperature, present temperature, and heater ON/OFF

status of the isothermal oven, LSV, and FPD, are indicated.

Figure 3.7

3.1.4

How to Check the Operating Status of the Detector

(1) The SYS tab with the detector to be checked is selected.
(2) The detector information can be checked. Refer to fi gure 3.6.

(3) When

is pressed, the following information are indicated

Input signal:

The analog signal from the detector, which has been sampled every 40
msec, converted to digital values, and then averaged by the given sample
rate

Filtered signal:

Input signal, which has been fi ltered using the fi lter constant set as a
detector signal parameter

2nd Edition : May 11, 2012-00