Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 182

Advertising
background image

<3. Basic Operation and Startup>

3-31

IM 11B08A01-01E

Figure 3.25

Calibration (validation) Method Setting Screen

(b) Calibration setting screen

(1) Press F4 key (Cal) on the Cal/Val Set (Main) screen. (The screen of Figure 3.26 appears.)
(2) Cal stream sets the calibration stream to be measured.
(3) Cal times sets the number of times of measurement of the calibration stream.
(4) Auto cal sets whether the calibration of this calibration number is conducted or not at

automatic calibration (effective only for the automatic calibration).

(5) ValStr# before Cal sets the number of the validation stream to be measured before

measuring the calibration stream.

(6) ValStr# after Cal sets the number of the validation stream to be measured after measuring

the calibration stream.

(7) Val before Cal times sets the number of times of measurement of the validation stream

before measuring the calibration stream.

(8) Val after Cal times sets the number of times of measurement of the validation stream after

measuring the calibration stream.