<5. EtherLCD>

5-92

IM 11B08A01-01E

When the stem identifying area is selected and the Set/Ent key is pressed, a strm identifying set
number is displayed in the second line below (from none up to Stream 31).
Use the up and down key to select the number and press the Set/Ent key, and the number is
refl ected.

• Setup for Distillation (setting possible per GCM) is as shown below.

(1) When distillation is set to Yes for more than one GCM, the SimDis Setup items on the

Table Menu (3/4) screen is shown.

(2) Distillation is added to the stream peak process (type) for the GCM with the Yes

distillation.

(3) Setting in the distillation screen is enabled for the GCM with the Yes distillation.

• When an over limit value is set as a GCM number, “Over limit” is displayed in the line below

and the set value is not refl ected.

• ATM valve selection is only possible for an ATM valve that can be set by GCM from oven

that belongs to SYS and which belongs to GCM. When an ATM valve that is set by another
GCM is specifi ed, the error message “Already set in another GCM” is displayed.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #
SYS1 to 6

○ *

None (default), Yes

ATM valve

number

○ *

None, 1-1, 1-2, 2-1, 2-2, 3-1, 3-2

* On the ATM valve setup

screen, only the number set

to Installed is selected and

displayed.

Strm

identifying

None, up to 3 streams, up to 7

streams, up to 15 streams, up to

31 streams

DO fi rst

number

for stream

identifi cation

(Slot number) − (relative D/O

number), (0), 1 to 5, When 0

is entered in slot number and

relative D/O number, 0 is set

in the database and asterisk is

displayed.

Stream valve

fi rst number

0 to total of stream valves, the

total stream valves are set on

the fundamental system settings

screen.

Number of

stream valves

0 to total of stream valves, the

total stream valves are set on

the fundamental system setting

screen.

It is necessary that stream valve

fi rst number + the number of

stream valves is less than or

equal to the total number of

stream valves.

Distillation

○ *

NA, Yes

*:

The settings are valid after rebooting or turning off/on its power.

(Example)
GCM 1 = up to 7 streams, stream identifi cation read signal of 1-bit, 3-bit (111) D/O fi rst number 1
for stream identifi cation
DO 1: GCM 1 stream identifi cation read signal
DO 2 to 4: GCM 1 stream identifi cation signal

GCM 2 = up to 15 streams, stream identifi cation read signal of 1-bit, 4-bit (1111) D/O fi rst number
1 for stream identifi cation
DO 1: GCM 1 stream Identifi cation read signal
DO 2 to 5: GCM 1 stream identifi cation signal

2nd Edition : May 11, 2012-00