Caution – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 138

Advertising
background image

<2. Installation, Piping, and Wiring>

2-21

IM 11B08A01-01E

(1)

24±1

O-ring

*: Provided by user

Case of electronics section
(Pressurized enclosure)

Insulated wires *

Unit: mm

Sealing dam *

Sealing compound *

17 or more

Gasket

Close-up plug

Body

Nut

Nipple
(connection threads)

Nipple

Nut

O-ring

Close-up plug

Body

Hex. key wrench *
(width across flats: 10 mm)

To the connection port of
the electronics section

Thread type and size: G3/4

Figure 2.15

Mounting procedure for sealing fi tting (accessory)

CAUTION

The cable end should be sealed in order to apply pressure to the electronics section. Otherwise,
power does not be supplied to the electronics section.

For the 3/4NPT connection port, use the connectors in the accessory kit to all of the connections.

TIP

Refer to “USERS’ GUIDELINES for Electrical Installations for Explosive Gas Atmospheres in General Industry”
for more details.

2nd Edition : May 11, 2012-00

2