Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 86

Advertising
background image

<1. Overview>

1-59

IM 11B08A01-01E

Run

Stop

Stop

Run

W7

W7

W8

S8

W1

W2

S1

S7

S7

Time

Stream sequence 1

Calibration 1

(Replaceable)

(Replaceable)

Stream sequence 1

Assinning calibration 1

Start of calibration (or validation)

End of calibration (or validation)

● In the case where a (manual) calibration/validation command is received during the stop

mode in the stream sequence status

The measurement status immediately changes to “Replaceable” in the stop mode in the

requested calibration (or validation) status.

Issue a command to start calibration (or validation) after confi rming that manual

replacement of the fl ow with the sample has been completed. When calibration (or
validation) is completed for all the streams specifi ed for the selected calibration (or
validation) number, the operation mode returns to the stop mode in the stream sequence
status.

Stop

Stop

Run

W7

W7

W8

S8

S7

S7

Time

Stream sequence 1

Calibration 1

(Replaceable)

(Replaceable) Stream sequence 1

Assinning calibration 1 Calibration (or validation) start command

Calibration (or validation) stop command

● In the case where a (manual) calibration/validation command is received during the pause

mode in the stream sequence status

The measurement status immediately changes to “Replaceable” in the stop mode in the

requested calibration (or validation) status.

Issue a command to start calibration (or validation) after confi rming that manual

replacement of the fl ow with the sample has been completed. When calibration (or
validation) is completed for all the streams specifi ed for the selected calibration (or
validation) number, the operation mode changes to the stop mode in the stream sequence
status.

Pause

Stop

Stop

Run

W7

W7

W8

S8

S7

S7

Time

Stream sequence 1

Calibration 1

(Replaceable)

(Replaceable) Stream sequence 1

Assinning calibration 1 Calibration (or validation) start command

Calibration (or validation) stop command

● In the case where a (manual) calibration/validation command is received during the run

mode in the stream specifi cation status

The measurement status is immediately prepared to be changed to the calibration (or

validation) status.

In the case of the stream specifi cation status (for continuous measurement), the

measurement status immediately changes to “Replaceable” in the stop mode in the
requested calibration (or validation) status.

In the case of the stream specifi cation status (for N times of measurement), the

measurement status changes to “Replaceable” in the stop mode in the requested calibration
(or validation) status when the Nth round of the measurement of the currently-measured
stream is completed.

2nd Edition : May 11, 2012-00