<5. EtherLCD>

5-44

IM 11B08A01-01E

• The initial screen displays the smallest oven number among those assigned to isothermal

ovens on the Temperature Control Set.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Oven #

Oven number

1

3

Vent #

Vent number

1

6

Carrier gas

number

Carrier gas number

---

1-1

1-2

2-1

2-2

3-1

3-2

* --- indicates not used.

The value is set on the Detector

Set screen.

Vent type

Vent type

---

BF

FF

Split

* --- indicates not used.

Vent fl ow rate Vent fl ow rate

Alpanumeric: 10 characters

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Operating Parameters

(5/8)

>

- Pressure unit kPa

- Vent fl owrate unit ml/min
- Sample volume unit ml

Menu

F1

F2

F3

F4

F5

F6

Figure 5.57

Example of the operating parameters screen (5/8)

F1 (Menu):

Displays the Table Menu screen.

TIP

If the pressure unit is changed, the gas pressure value including the one set to EPC controlled will be converted
according to the new unit. For example, if the unit is changed from kPa (the value range is set to 0 to 500 kPa) to
psi (the value range 0 to 72.5 psi), 250 kPa is converted to and displayed as 36.25 psi.

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Pressure unit

kPa

psi (see the defi nition of SI

units.)

Vent fl ow rate

unit

ml/min

Sample

volume unit

ml

μl

2nd Edition : May 11, 2012-00