<Appendix D Action of External Input and Output Signals>

App.D-19

IM 11B08A01-01E

The Calibration/Validation function of D/I is available for semiautomatic calibration. To enable the

function, the relevant calibration or validation number is required to be specifi ed in advance.

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

D/I Setup

Slot #

1 DI # 1

>

- Signal name G1-CAL1

- Process assign Cal/Val
- GCM number 1
- Cal/Val number Calibration 1

Menu

Status

Slot#

DI#

F1

F2

F3

F4

F5

F6

Figure 26 Process Specifi cation: Calibration/Validation (D/I Setting Screen on GC-HMI EtherLCD)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Cal/Val Setup (Main)

GCM # 1
> - Cal(Val) method Semi-auto
- Auto start date

2011/11/22

- Auto start time 17:00
- Time interval 001 day:00:05
Auto cal status Stopping

Menu

Start

Stop

Cal

Val

GCM#

F1

F2

F3

F4

F5

F6

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Calibration Setup

GCM #

1 Cal # 1

>

- Cal stream 4

- Cal times 3
- Auto cal Executed
- ValStr# before Cal 3
- ValStr# after Cal 4
- Val before Cal times 1
- Val after Cal times 1

Menu

Peak

Main

GCM#

Cal#

F1

F2

F3

F4

F5

F6

Figure 27 Calibration/Validation Setting Screen and Calibration Screen on GC-HMI EtherLCD.

Operation Mode

The Operation Mode function of D/I is a function to give a command to change the operation

mode to a specifi ed operation mode for a specifi ed GCM using contact input as a trigger.