2) hmi unit – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 369

Advertising
background image

<6. Maintenance>

6-16

IM 11B08A01-01E

9

F0609.ai

11

Figure 6.9

17

F0610.ai

14

15 12

18

16

13

Figure 6.10

21

F0611.ai

20

19

Figure 6.11

(2) HMI Unit

(1) Stop the operation and the supply of power.
(2) Open the door of the electronics section using a key included in the accessory kit.
(3) Loosen a screw of the HMI unit cover on the back of the door to remove the cover.

Remove three cable connectors attached to the HMI unit and one connector attached to the
status board.

(4) Remove eight screws around the HMI unit, and remove it with the bracket.
Confi rm that the rubber is securely fi tted in the groove on the door.

2nd Edition : May 11, 2012-00