User program edit screen – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 278

Advertising
background image

<5. EtherLCD>

5-27

IM 11B08A01-01E

F3 (Edit):

Displays the User Program Edit screen.

F5 (Stop):

Executes the forced stop command.

• After selecting a SYS number, the option displayed is Enable if the execution status is

currently disabled, and Disable if it is currently enabled.

• Pressing F5 (Stop) displays Input script number (SYS #). Entering the target script number

for the end of peak detect executes the forced stop command.

TIP

The Status is refreshed automatically at a fi xed interval (1 second).

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Enable

Disable

Enable

Status

Stopping

Executing

User Program Edit screen

Press F3 (Edit).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

User Program Edit

Type #1 NO #1
- AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
- AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
- AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
- AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
- AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
- AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
- AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
- AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA
> - AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA

Main

Save

Add

Del

Type#

NO#

F1

F2

F3

F4

F5

F6

Figure 5.32

Example of the user program edit screen

F1 (Menu):

Displays the Operation Menu screen.

F2 (Save):

Saves the script fi le.

F3 (Add):

Add a command line.

F4 (Del):

Delete a command line.

F5 (Type #): Specifi es a type number.
F6 (No #):

Specifi es a SYS number for the end of peak detect script type; or a period script
number for the period script type.

TIP

• For this operation, the user level C+ is required.
• The following values can be entered:

0 to 9, A to Z, ( ) * / + – =, space and backspace.

• Type

number

1: End of peak detect script
2: Period script