<6. Maintenance>

6-34

IM 11B08A01-01E

CAUTION

The isothermal oven is extremely hot after turning off the power immediately. Keep the purging
air supplied for more than an hour after turning off the power. Keep hands away from the oven
components.

(3) Turn off the supplies of the purging air, carrier gas, hydrogen (or nitrogen) gas for

combustion, combustion air, and sample gas.

(4) Open the control unit door using a key included in the accessory kit.
(5) Remove the two couplings connected to the methanizer. When removing the couplings,

label them for ease of identifying the connections later.

(6) Remove the two wires connected to the terminals of the methanizer.
(7) Remove the two screws to remove the methanizer.
(8) Install a new methanizer.
(9) Reconnect the wires and pipes in the reverse order of steps (4) to (7). Confi rm wiring and

perform a leak test.

<Replacement Parts>

K9192TG

F0230.ai

Heater terminals

Screw

Screw

Outlet

Inlet

Figure 6.30

Methanizer

(6) Rotary Valve and Valve Seat

The rotary valve located in the isothermal oven is used as a sampling valve and a column
switching valve. The valve has two types: general type and high temp. type. The specifi cations
for the rotary vale used are listed in the parts list in the operation data. Check against the code
chart (Table 6.2) and replace as required.

Table 6.2

Model

Suffi x code

Option code

Description

GCRV

•••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• Rotary valve

Bellows

material

-G

-M

•••••••••••••••••

•••••••••••••••••

Rubber

Metal

O-Ring

material

B

K

•••••••••••••••••

•••••••••••••••••

Viton

Kalrez

2nd Edition : May 11, 2012-00