Calibration peak set – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 326

Advertising
background image

<5. EtherLCD>

5-75

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #
Cal #

Cal Number

0

6

Cal stream

○ 0

31*

* Only stream numbers and cal

streams belonging to GCM can

be set. (Stream setup screen )

Cal times

○ 0

30

Cal times + Val before cal times

+ Val after cal times must not

exceed 31.

Auto cal

specifi cation

Not done, Done

ValStr #

before and

after cal

○ 0

31**

** Only stream numbers and Val

streams belonging to GCM can

be set. (Stream setup screen)

Val before

and after cal

times

○ 0

30

Cal times + Val before cal times

+ Val after cal times must not

exceed 31.

Calibration peak set

Press F2 (Peak).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Calibration Peak Setup

Peak # 1
>

- Peak name

- Execution Executed
- Measuring unit 1 ( )
- Measuring range 1.000
- Std conc

1.000

- Std area 950.000
- Calib factor 2.063

Cal

Peak#

F1

F2

F3

F4

F5

F6

Figure 5.83

Example of calibration peak set screen

F2 (Cal):

Displays the calibration set screen

F6 (Peak #): Specifi es a peak number.

• Displays the stream peak set in Cal stream.

2nd Edition : May 11, 2012-00