C) gcm operating status display area – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 215

Advertising
background image

<4. GC-HMI (touch panel)>

4-28

IM 11B08A01-01E

(C) GCM operating status display area

Buttons for moving to a dialog for changing GCM settings

F0433.ai

Buttons for changing GCM operating status/mode

GCM detector/valve control buttons

GCM control buttons

Figure 4.33

Example of GCM Operating Status Display Area

(1) Changing the operating status

This section describes how to change the operating status (GCM status/operating mode/
measurement status) of the active GCM or of each SYS.

GCM status

Select the GCM status Process or Manual.

Process:

Status for performing measurement and calibration.

Manual:

Status for performing manual operation.

TIP

This operation is available if the user level is set to C or higher.
Status change is only possible when the operating mode is Stop.

Operating mode

Select the desired operating mode of the active GCM or of each SYS from the following:

Run:

Runs

measurement.

Pause:

Pauses the run.

Stop:

Stops

the

run.

Cancel command:

Cancels the operating mode command.

Forced stop:

Forces operating mode to stop.

TIP

• The Run, Pause, Stop, and Cancel command actions are enabled for user level B or higher.
• The Forced stop action is enabled for user level C+.
• The Pause action is enabled only for the active GCM. It is disabled for each SYS.
• The operating mode can be set on the analyzer overview screen and the analyzer operation screen. The

Forced stop action can only be executed on the analyzer operation screen.

2nd Edition : May 11, 2012-00